Chapter III · THERMO-ELECTRIC PYROMETERS (continued 2)
Reading edition
Owing to the magnitude of the error caused by changes in the cold junction, the thermo-electric method is not suited to the measurement of atmospheric temperatures, or for explosive magazines or cold stores. In such cases instruments of the resistance type, to be described later, are used.
Temperature of Steam, Exhaust Gases, etc.—For measuring the temperature of ordinary or superheated steam, the exhaust gases from internal combustion engines, etc., iron-constantan junctions, with suitable indicators, are satisfactory. When placed in a pipe the junction should be as nearly as possible in the centre, so as to avoid the cooling effect of the walls. Several junctions, situated in different parts of the pipe, may be used with a single indicator and suitable switchboard. The above remarks also apply to the hot-blast for blast furnaces, and similar instances where the temperature does not exceed 900° C.
Measurement of Differences of Temperature.—Cases frequently arise in practice in which the difference in temperature between two points is required, and if this difference be subject to rapid changes, a mercury thermometer, from its large mass, would not respond with sufficient rapidity to indicate these changes. In such cases a circuit is made after the manner of fig. 2, one junction being located at each point; thin wires of iron and constantan being used. For small differences—1° C. or less—a mirror galvanometer should be used. Calibration may be performed by placing one junction in hot water and the other in cold, the water temperatures being read with an accurate thermometer.
Advantages of the Thermo-electric Method of Measuring Temperatures.—Compared with other methods, the thermo-electric possesses the following points of superiority:—(1) Simplicity, no special experiment being necessary to obtain a reading; (2) cheapness of outfit; (3) adaptability to a variety of purposes; (4) ease of repair in case of damage; (5) robustness, not being liable to get out of order under workshop conditions; and (6) suitability to the purpose of a centrally controlled installation. The drawbacks are:—(1) Liability to error owing to fluctuations in the cold junction (which may be avoided with care); and (2) lack of sensitiveness at very high temperatures compared with the resistance method—a point seldom of great practical importance, as the limit of accuracy is usually within the amount by which an ordinary furnace fluctuates in temperature under working conditions.